How to detect and determine the balance machine and balancing accuracy

 

I. Common anomalies of the fault diagnosis

 

1, most of the balancing machine measuring system has a "self-test" function, this function can detect the measuring system itself is normal. And "self-test" function corresponds to the operating buttons may be labeled as "self" or "TEST" and so on. Users can balance the machine normally, it will support a variety of setting conditions (bearing means and a, b, c, and two radii, etc.) fixed, then the measurement system into the "self" state, corresponding to the boundary conditions recorded under the "self" state of the display reading. When the operator that the measurement system has a problem, you can restore your system to the measurement corresponding to the original set of supporting conditions of the "self" status, then check the measuring system display readings are normal.

2, balance machine shows the angle of unbalance the total at roughly 180 degrees or 0 degrees. First, to confirm the normal operation of measuring and balancing machine is still some residual rotor unbalance (or even weight on both the rotor surface, respectively, to install two different phase unbalance), measured in the normal rotation of the rotor case: pull 1, the sensor under the cord, to see whether the meter display value changes. If significant change is that the sensor cable and sensor everything is normal. If no change, then prove that this sensor lines or sensor problems. No. 1 with the plug inserted, and then unplug the 2 wire sensor, the same method can determine the sensor line and sensor 2 is normal. Users can find other professionals in the sensor line and sensor control on a faulty sensor or sensor wire repair.

3, the balance of the residual unbalance in a larger machine, the fault is not obvious. However, small residual unbalance, balancing machine to enter the time you start measuring, the display value of the angle of the total balance is changing. Sometimes the angle changes within a certain range, and sometimes 360-degree angle in the range of variation.

 In order to reduce the same frequency, frequency, frequency interference, the workpiece supporting roller bearing shaft diameter should be avoided and the entire fraction or multiple of the same or close to, in order to avoid interference. For example: wheel diameter of 101 mm, so it is best to avoid the use of 91 - 111,46 - 55,32 - 36 mm range of the shaft bearing.

 Check the rotor assembly part of the strict stability. Axis as the use of technology should be focused on checking with the shaft and hole.  Check the rotor axis (process axis) and the state of contact at the wheel, if the shaft roughness, knife or roller surface pattern was injured, will lead to small signal instability.

 Check the wheel and the rotor axis (process axis) contact at the state, if the wheel contact surface of a continuous smooth outer surface has been damaged, can lead to small-signal instability. When the roller contact surface have obvious injuries, the balance amount of the display is very unstable.

 state of the cleaning roller and stop the rotor axis (process axis) of the contact surface, add appropriate amount of lubricant.

 careful calculations, if your balance is too high accuracy.

4, the balance of the residual unbalance in a larger machine, the fault is not obvious. However, the residual unbalance is small, with the balance after each restart measurements, showing the angle value is always unbalanced roughly 180-degree change in the way.

① The balance amount with the high side.

② Check the rotor assembly part of the strict stability. Use of technology such as axes, you should check with strict emphasis on the gap axis and the hole, ovality and taper errors, especially in ovality and taper error.

Two. Four-point method Troubleshooting

Detected by the above, maintenance is completed, if the fault still exists, the balance can be described by the following four o'clock machine balancing machine test method to test. Under normal circumstances, so the user can use the physical rotor. If necessary, users can use the following process and record all the data then available to balance the machine manufacturer, the manufacturer can determine the basis of this record more than 90% of the fault.

1, the first rotor balance to be able to achieve the highest accuracy, and then prepare a residual rotor unbalance is greater than 30 to 50 times the removable weights (according to the following procedure: Using weights 100 g, the residual amount of the rotor should be less than 2 to 3 grams). Rotor and load balancing machine coupling and the form of the same, directly following the testing process.

2, check the balance of the various machine settings are normal.

3, either side of the rotor circumference at the specified weight add 45 degrees to a known weight of removable weights (the weights by assuming 100 g).

Start balancing machines, balancing machines indicated value corresponding to the surface should be 40-50 degrees, 90-110 g (90% according to weight ratio).

4, shutdown will be removed with weights installed on the same side of the circumference of the rotor angle of 135 degrees.

Start balancing machines, balancing machines indicated value corresponding to the surface should be 130-140 degrees, 90-110 g (90% according to weight ratio).

5, stop and then remove the weights installed on the same side of the circumference of the rotor angle at 225 degrees.

Start balancing machines, balancing machines indicated value corresponding to the surface should be 220-230 degrees, 90-110 g (90% according to weight ratio).

6, stop and then remove the weights installed on the same side of the circumference of the rotor angle at 315 degrees.

Start balancing machines, balancing machines indicated value corresponding to the surface should be 310-320 degrees, 90-110 g (90% according to weight ratio).

7, the same process, the other side of the rotor weight specified point on the circumference at four weights in order to install this check.

8, such as balancing machines can be up to the requirements stated above is considered to balance the machine without any problems.

9, such as balancing machines can approximate to the above requirements is considered minor balancing machine calibration inaccuracies. This problem usually caused by a decline in operating efficiency, without causing large errors.

If there was a slight balancing machine calibration inaccuracies, the user can balance the machine matching the content rule out the instructions, may also notify the manufacturer for processing.

10, such as balancing machine does not meet the above requirements, and the error is very large, the user shall notify the manufacturer for processing.

11, special circumstances, the user to balance the machine accuracy or quality of doubt, you can balance the machine on the four-point method for fault diagnosis.

Sometimes users will find that the rotor has been balanced on the balancing machine precision, the above method of testing that balancing machines are all normal, but the actual installation of the rotor running, or have some degree of vibration, then you need to carefully analyze other aspects of the process why.

Three. Careful analysis of the vibration of the reasons, the right to determine the accuracy of the real balance

Mechanical vibration caused by many reasons, including its balance of rotating parts caused by the vibration of the most direct. But this does not mean that the rotor balancing machine balance in a good future, they can solve all vibration problems.

Rotor imbalance is due to the rotor's center of gravity, the rotation of the rotor geometry deviation from the centerline. The rotor rotates the resulting centrifugal force together with the rotor rotation, the rotor and the rotor in order to stimulate the installed base (such as: bearing, frame, chassis, body, etc.) to vibrate.

The following will discuss some other related process issues.

1, the process itself, problems balancing machine

Precision balance, should be considered universal coupling, connectors and other related key influence. Small rotor balancing process requires additional ancillary quality of the cursor, but also need to consider their impact on the accuracy of the final balance. When these connections related to the amount of close to or even caused by the imbalance of the rotor balance is greater than the precision of user requirements, the balance machine shows the unbalance is not reliable. Drive with a universal coupling for the balancing machine, universal coupling should also be regularly checked the balance of accuracy.

When using the balance shaft technology, process shaft runout accuracy can not be greater than the rotor balance index e (eccentricity), or 1800 high-precision balance must be flipped. Or process axis scrapped.

In addition, the process structure and the rotor shaft axis of the actual installation of the structure, should be made into the same or similar to ensure that its mass distribution consistent with the actual installation of the state.

2, the rotor of the process issues

Balancing machine using the user should try to install the real state of rotor balance, and ensure its rigidity.

Axis to balance the use of technology, there is the actual installation process and the rotor shaft axis error, such as the rotor shaft in the process and the actual axis of different installation methods, keyway structure, process and the actual shaft axis and the beating of their own, these factors will lead to uneven measurable change.

Axis to balance the use of technology, we must ensure that the rotor mounting hole dimensional accuracy and geometrical tolerances, such as: the gap is large, oval or tapered hole error, low roughness is absolutely not allowed.

Another must ensure that the rigidity of the rotor. If an internal sliding splined rotor, active block, swing knife block and other similar structures, such as the rotor is never reach the ideal balance of accuracy.

3, the actual installation process issues rotor

Rotor balancing machine has reached the required balance of precision, the actual installation of the rotor operation, if the actual installation process do not seriously consider certain factors, or the changes will result in a balance accuracy, as well as causing a certain degree of vibration.

The first is the actual installation of shaft runout, this factor is very important. Because the rotor imbalance is due to the rotor center of gravity, deviation from the geometric center of rotor rotation. The rotor rotates the resulting centrifugal force together with the rotor rotation, the rotor and the rotor in order to stimulate the installed base of vibration. The balance of good eccentric rotor is installed in the actual installation of the shaft has, of course, will lead to the vibration of the rotor and the machine.

In addition, coaxial with the rotor will cause the rotor and other parts of machine vibration, such as belt pulleys, couplings, rotor, etc., their precision will balance together with the rotor in the machine vibration.

4, other issues

Our technical staff during the service found that there are some balancing machine users in their production process, pay attention to grasp the early stage of manufacturing quality of its products, and operational performance of its rotor completely dependent on the balancing process, it is very not serious. For example:

Product design process does not consider the process of balancing;

Products based components (such as brackets, etc.) cut corners or even seriously flawed;

Fan rotor or rotor shape welded face beating too large;

Uneven distribution of the quality of the rotor casting error is too large;

The rotor mounting hole precision and roughness is too low;

The internal structure of the rotor there is a serious non-rigid factors (such as: weld, loose rivets, etc.);

Roughness and accuracy of technology-axis low;

Actual installation of the rotor shaft is too low precision and roughness;

All of these factors will lead to reduced accuracy even balance balance failure.

In addition, dynamic balancing machine is a type test machine instrumentation products. To balance the machine's rough handling (such as: welding on the balancing machine, sanding, etc.) will also directly affect machine performance and life balance, leading to balanced failure, do pay attention.

How to detect and determine the balance machine and balancing accuracy

What is dynamic balancing

How to make a reasonable balancing accuracy

 

   Our balanced electromechanical tester uses industrial single-chip technology and has a fast start-up speed, making it especially suitable for factory work environments! The technology that can be measured after turning on, memorizing and stopping, can input and change the automatic calculation result of the parameter, the electrical measurement box has the data record of the measurement result, and the measurement data sd card records the function.

Our hard-bearing balancing machine is a permanently calibrated hard-support balancing machine.

  The rotors that balance different shapes and weights do not need to be recalibrated. The size and angle of the unbalanced values can be measured at a single start.

 

 